Hi,
I'm trying to calibrate my 1044_1 PhidgetSpatial Precision 3/3/3 IMU, with the newest firmware (v. 502).
However, each time I perform a calibration in 2-Axis mode, I can see the red circle as I rotate the IMU around the Z-axis, but as soon as I hit the STOP button, the window crash, and I don't have any value or successful message with the calibration parameters.
I tried to downgrade the firmware version to the previous one (500), and I have been able to perform the calibration and obtain calibration values, as well as the green circle showing the calibrated results.
It looks like a bug in the newest firmware (502). Would it be possible to fix it?
Best